Features
The UVMINILOG is designed for high performance
and reliable logging of ultraviolet radiation.
It uses sophisticated microcontroller technology and
low noise SiC based UV detectors allow up to 18 months of permanent measurement
and logging without battery charging.
Applications include dose monitoring of UV sensible
goods such as artworks or compound materials. Other fields are dose monitoring
of UV hardening systems or sun UV (e.g. Erythema or UVA + UVB) monitoring of
persons, animals or plants.
Sensors
UV Sensors included (available with one or two UV
sensors)
The UVMINILOG includes sensors from the sglux
TOCON series. The TOCON devices use modern hybrid
technology to cancel unwanted signal disturbances caused by moisture or
electromagnetic radiation. All TOCONs use a silicon carbide (SiC) detector
chip.
By default we ship the device with a UVA-UVB-UVC
broadband photodetector which is NIST traceably calibrated with a Xenon high
pressure lamp (Newport 6251). Other NIST calibrated sensors such as sun
Erythema, UVC disinfection, UV hardening and low UV sensing sensors are
available on request. Please contact us for further assistance.
Optional Sensors
The UVMINILOG can be equipped with four further sensors:
| |
Measure: |
Working range: |
Accuracy |
 |
Temperature |
-10°C to +58°C |
±0.1°C (5°C to 45°C) ±0.2°C
(-10°C to +58°C) |
 |
Relative Humidity |
0-100% rel. Hum.( -20°C to +65°C) |
±2% rel. hum. (10-85% rel. hum., 0 to 40°C)
±4% rel. hum. (85-95% rel. hum., 0 °C to 40°C) |
 |
Pressure |
0-2500 mbar abs. |
±2.5 mbar (750-1100 mbar absolute) |
 |
Acceleration |
±10 G / ±2 G sel. |
±0.15 g (25 °C) |
Device
Data Transfer and Software
With the provided software Setup the user customises
the properties of the UVMINILOG. With the software Reader the USB data
transfer is started. The Viewer is used for graphical displaying. The data can
be exported as a csv file for analysing in standard softwares like Excel or
Origin. The software Online is displaying online measurements.
Logging Features
Record limits can be set for all used sensors. If a
signal exceeds a limit the red LED will flash.
Measurements can be started via a connected computer
(date and time for the start can be chosen) or manually by a push-button.
The push-button can be used as a marker to mark special
events in the recording.
For each of the possible six sensors a measurement rate
between 1s and 12h can be chosen.
Prediction feature calculates memory and battery
capacity for the chosen measurement rates.
For monitoring of sensitive transport goods a shock
measurement can be activated (if acceleration sensor is equipped). Therefore a
threshold can be chosen. Every acceleration above this threshold is recorded.
The mixing gravitational acceleration is not taken into account.
Attachment
The UVMINILOG has been developed for monitoring
of moving goods and beings. It can be attached with a belt or three screws.
Versions
The UVMINILOG is available in different
configurations. Besides the choice of sensitivity and spectral response of the
UV sensors and the choices of additional sensors the device can be delivered in
a waterproof (IP 67) housing. Please contact us for customised specifications
and housings.
Specifications of the UVMINILOG
|
Sensors and Output |
 |
Number of SiC UV sensors: 1 - 2 |
 |
Specs. of UV Sensor: please have look at the datasheet with
detailed specifications of the used sensor. |
 |
Working range and accuracy of optional Sensors: see
above |
 |
Storage rates: |
| |
min. storage rate UV Intensity: 2/day |
| |
max. storage rate UV Intensity: 50/second |
| |
min. storage rate Temperature: 2/day |
| |
max. storage rate Temperature: 1/second |
| |
min. storage rate rel. humidity: 2/day |
| |
max. storage rate rel. humidity: 1/second |
| |
min. storage rate pressure: 2/day |
| |
max. storage rate pressure: 10/second |
| |
min. storage rate acceleration: 2/day |
| |
max. storage rate acceleration: 50/second |
| |
|
|
Standard Parameters of the housing (varies with
required features) |
 |
Dimensions (BxHXD): 78 x 39 x 24 mm |
 |
Weight: 60 g |
| |
|
|
Additional technical data |
 |
Operating temperature: -15 to +65 °C |
 |
Storage temperature -20 to +70 °C |
 |
Capacity lithium-polymer battery: 900 mAh |
 |
Data storage: >2,000,000 parameters |
|